IEECAS OpenIR  > 加速器质谱中心
Trace determination of selenium in biological samples by CH4-Ar mixed gas plasma DRC-ICP-MS
Guo, W (Guo, Wei)[ 1 ]; Hu, SH (Hu, Shenghong)[ 1 ]; Wang, YX (Wang, Yanxin)[ 1 ]; Zhang, LY (Zhang, Luyuan)[ 3,4 ]; Hu, ZC (Hu, Zhaochu)[ 2 ]; Zhang, JY (Zhang, Jiangyi)[ 5 ]
2013-05-30
发表期刊MICROCHEMICAL JOURNAL
卷号108页码:116-122
文章类型期刊论文
摘要

Direct determination of ng g(-1) level of selenium by ICP-MS is complicated because of the presence of serious spectral interferences, and the high first ionization energy of Se (9.75 eV) also results in low analytical sensitivity. In this study, a reliable method for accurate determination of Se in biological samples using CH4-Ar mixed plasma-DRC-MS was evaluated. The predominant interfering Ar-40(2)+ (from ion source) on the most abundant Se-80 isotope was successfully eliminated by the charge transfer reaction using methane as DRC gas. The minor interferences of oxide or hydroxides ions (Zn-64 O-16(+), (NiO+)-Ni-64-O-16 and (CuOH+)-Cu-63-O-16) and double charged ions (Tb-159(2+), Dy-160,161(2+) and Gd-160(2+)) originating from sample matrix were decreased down to 50% using CH4-Ar mixed gas plasma instead of the conversional pure Ar-ICP. Meanwhile, the poor sensitivity of Se+ was improved by a factor of 3 due to the carbon-enhancement effect in CH4-Ar mixed gas plasma. In addition, to avoid the interference (BrH+)-Br-79 on Se-80(+), an elaborate evaporation procedure was used to remove the Br from the digestion solution. The proposed method was applied to direct determination of Se in fifty biologically related SRMs. Our results showed that the values of most SRMs were found to agree well with the certified or recommended values. This method has great potential for the determination of trace or trace level of Se in various biological samples.

关键词Selenium Biological Samples Ch4-ar Mixed Gas Plasma Drc-icp-ms
DOI10.1016/j.microc.2012.10.006
收录类别SCI
语种英语
引用统计
被引频次:40[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.ieecas.cn/handle/361006/9913
专题加速器质谱中心
通讯作者Guo, W (Guo, Wei)[ 1 ]
作者单位1.State Key Laboratory of Biogeology and Environmental Geology, School of Environmental Studies, China University of Geosciences, Wuhan, 430074, PR China;
2.State Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences, Wuhan, 430074, PR China;
3.Xi'an AMS center and State Key Laboratory of Loess and Quaternary Geology, Institute of Earth Environment, Chinese Academy of Sciences, Xi'an, 710075, PR China;
4.Engineering Geology and Water Resources Department, Institute of Geology and Geophysics, Chinese Academy of Sciences, Beijing 100029, PR China
推荐引用方式
GB/T 7714
Guo, W ,Hu, SH ,Wang, YX ,et al. Trace determination of selenium in biological samples by CH4-Ar mixed gas plasma DRC-ICP-MS[J]. MICROCHEMICAL JOURNAL,2013,108:116-122.
APA Guo, W ,Hu, SH ,Wang, YX ,Zhang, LY ,Hu, ZC ,&Zhang, JY .(2013).Trace determination of selenium in biological samples by CH4-Ar mixed gas plasma DRC-ICP-MS.MICROCHEMICAL JOURNAL,108,116-122.
MLA Guo, W ,et al."Trace determination of selenium in biological samples by CH4-Ar mixed gas plasma DRC-ICP-MS".MICROCHEMICAL JOURNAL 108(2013):116-122.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Trace determination (535KB)期刊论文作者接受稿开放获取CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Guo, W (Guo, Wei)[ 1 ]]的文章
[Hu, SH (Hu, Shenghong)[ 1 ]]的文章
[Wang, YX (Wang, Yanxin)[ 1 ]]的文章
百度学术
百度学术中相似的文章
[Guo, W (Guo, Wei)[ 1 ]]的文章
[Hu, SH (Hu, Shenghong)[ 1 ]]的文章
[Wang, YX (Wang, Yanxin)[ 1 ]]的文章
必应学术
必应学术中相似的文章
[Guo, W (Guo, Wei)[ 1 ]]的文章
[Hu, SH (Hu, Shenghong)[ 1 ]]的文章
[Wang, YX (Wang, Yanxin)[ 1 ]]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。