Performance of Accelerator Mass Spectrometry for I-129 using AgI-AgCl carrier-free coprecipitation
Zhang, LY (Zhang, Luyuan)[ 1,2 ]; Hou, XL (Hou, Xiaolin)[ 1,2 ]; Zhou, WJ (Zhou, Weijian)[ 1,2,4 ]; Chen, N (Chen, Ning)[ 1,2 ]; Liu, Q (Liu, Qi)[ 1,2,3 ]; Luo, MY (Luo, Maoyi)[ 2,4 ]; Fan, YK (Fan, Yukun)[ 1,2,3 ]; Fu, YC (Fu, Yunchong)[ 1,2 ]
2013-01-30
Source PublicationNUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS
Volume294Pages:276-280
Subtype期刊论文
Abstract

I-129 has been successfully applied as tracer in environmental, geological, and oceanographic research. For samples with low stable iodine concentration and ultra low level I-129, the sample preparation technique to separate iodine prior to AMS measurement has been a bottleneck, limiting the applicability of I-129 We have reported a carrier-free method, using coprecipitation, to avoid the potential introduction of I-129 through the use of stable iodine carrier iodine. In this work, the detection limit and the analytical uncertainty of this method are investigated and minimum sample amount required to obtain reliable analytical results are estimated. The method is validated with a series of samples in ranges of known iodine concentrations and I-129/I-127 ratios. The results confirm our previous conclusion that an AMS target containing 5.0 mu g iodine can be used for analyzing samples with I-129/I-127 > 10(-12), and that for samples with I-129/I-127 < 10(-13) more than 25 mu g iodine is necessary.

KeywordIodine-129 Carrier Free Coprecipitation Ams Minimum Sample Amount Detection Limit
DOI10.1016/j.nimb.2012.06.023
Indexed BySCI ; EI
Language英语
Citation statistics
Document Type期刊论文
Identifierhttp://ir.ieecas.cn/handle/361006/10051
Collection加速器质谱中心
Corresponding AuthorHou, XL (Hou, Xiaolin)[ 1,2 ]
Affiliation1.SKLLQG, Institute of Earth Environment, China Academy of Sciences, Xi’an 710075, China;
2.Xi’an AMS Center, Xi’an 710061, China;
3.Graduate University of the Chinese Academy of Science, Beijing 100049, China;
4.Xi’an JiaoTong University, Xi’an 710049, China
Recommended Citation
GB/T 7714
Zhang, LY ,Hou, XL ,Zhou, WJ ,et al. Performance of Accelerator Mass Spectrometry for I-129 using AgI-AgCl carrier-free coprecipitation[J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS,2013,294:276-280.
APA Zhang, LY .,Hou, XL .,Zhou, WJ .,Chen, N .,Liu, Q .,...&Fu, YC .(2013).Performance of Accelerator Mass Spectrometry for I-129 using AgI-AgCl carrier-free coprecipitation.NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS,294,276-280.
MLA Zhang, LY ,et al."Performance of Accelerator Mass Spectrometry for I-129 using AgI-AgCl carrier-free coprecipitation".NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS 294(2013):276-280.
Files in This Item:
File Name/Size DocType Version Access License
Performance of Accel(575KB)期刊论文作者接受稿开放获取CC BY-NC-SAApplication Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Zhang, LY (Zhang, Luyuan)[ 1,2 ]]'s Articles
[Hou, XL (Hou, Xiaolin)[ 1,2 ]]'s Articles
[Zhou, WJ (Zhou, Weijian)[ 1,2,4 ]]'s Articles
Baidu academic
Similar articles in Baidu academic
[Zhang, LY (Zhang, Luyuan)[ 1,2 ]]'s Articles
[Hou, XL (Hou, Xiaolin)[ 1,2 ]]'s Articles
[Zhou, WJ (Zhou, Weijian)[ 1,2,4 ]]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Zhang, LY (Zhang, Luyuan)[ 1,2 ]]'s Articles
[Hou, XL (Hou, Xiaolin)[ 1,2 ]]'s Articles
[Zhou, WJ (Zhou, Weijian)[ 1,2,4 ]]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.